1-Non staining count&Trypan blue staining count
Non staining counting (invention patent number: ZL 2013101014030): Without staining, cells can be directly counted and differentiated between live and dead cells, suitable for cultured cells in good condition and with a clean background; Trypan blue staining counting: a classic method for distinguishing between live and dead cells, with more intuitive color changes compared to non staining methods.

2-Hemocytometer&Manual comparison
The instrument is designed according to the "Blood Cell Counting Board" counting standard (JJG 552-1988 Blood Cell Counting Board Trial Verification Regulations) and can use standard blood cell counting boards; Cell concentration=(sum of four large square cell numbers)/4 x (dilution ratio) x 104,The accuracy of counting results can be manually verified.

3-Fully automatic intelligent linkage system&Cell count slide
Adaptive intelligent autofocus technology can adaptively adjust the focal length (Z-axis scanning), ensuring clear and sharp cell images in different regions and states. Combined with precise X and Y axis motion, it can achieve multi area image acquisition (up to 21 target surfaces); The cell counting slide adopts a two well design, with a 0.1mm counting cell height, fast cell sedimentation, fewer layers, and more accurate counting.

4-High definition imaging&Intelligent software recognition analysis
The instrument is equipped with 10X optical magnification and a 10 million color camera, making cell imaging clearer; Advanced image processing algorithms can automatically split clustered cells, ensuring high-precision image recognition. Customizable software recognition parameters can be used to re identify and analyze counting results.

5-Complies with FDA 21 CFR Part 11 and GMP requirements